• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; JCX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; JCX Back to Search Results
Catalog Number PMX110
Device Problems Device Operational Issue (2914); No Pressure (2994)
Patient Problem No Patient Involvement (2645)
Event Date 01/13/2016
Event Type  malfunction  
Manufacturer Narrative
Results: there was no visible damage to the exterior of the pump.Conclusions: evaluation of the returned device revealed that the pump was fully functional.During functional testing, the pump was plugged in, powered on, and power cycled multiple times.The pump was able to produce fully adjustable vacuum.The root cause of this complaint could not be determined.Penumbra pumps are 100% functionally tested during in-process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
During preparation for a thrombectomy procedure, the penumbra system aspiration pump max 110v (pump max) was powered on and the motor made a quiet noise; however, no vacuum was produced.The defective pump max was found prior to use and was not used for the procedure.The procedure continued using a new pump max.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer (Section G)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
kathleen kidd
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key5427507
MDR Text Key37982197
Report Number3005168196-2016-00156
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 01/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue NumberPMX110
Device Lot NumberF17857-31
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/13/2016
Initial Date FDA Received02/10/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-