| Catalog Number |
48824014 |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
01/17/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a revision surgery had to be performed because the plate got pulled out at c7.
|
| |
|
Additional Manufacturer Narrative
|
|
Catalog# 48824014.Method: device history review; complaint history review; risk assessment; results: the ifu specifically states "bone stock compromised by disease, infection or prior implantation which cannot provide adequate support and/or fixation to the devices" the stryker rep reported that the patient had poor bone quality and that this plate was put in where another implant was previously implanted.Conclusion: the probable root cause is likely related to the bone quality.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision surgery had to be performed because the plate got pulled out at c7.
|
| |
|