BOSTON SCIENTIFIC - COSTA RICA (COYOL) JAGTOME¿ RX 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
Model Number M00573040 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/19/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4) guidewire distal tip detached exposing the tip of the metal corewire.The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that there were a jagtome rx 44 was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2016.According to the complainant, during the preparation and outside the patient, it was noted that the device bowed out of plane and that its preloaded guidewire had malfunctioned.The ptfe coating of the preloaded guide wire was delaminated exposing the very tip of the metal corewire.The procedure was completed with another jagtome rx 44.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be "stable".
|
|