| Model Number |
1007607 |
| Medical Device Problem Codes |
Break (1069); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/19/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The patient is currently implanted with the rns system which is programmed for detection.Device history and sterilization review was performed: no evidence of factors that could have caused or contributed to the reported event.All sterilization results met requirements as specified in the specifications.Device history records for product assembly met requirements and passed quality control inspections.The depth lead and stylet were returned and investigated.Investigation identified that the style failed near where it is joined to the stylet retainer.The stylet failed due to a bending load.Neuropace is unable to identify the root cause of this failure.
|
| |
|
Event or Problem Description
|
|
During an rns system implant procedure and after the surgeon placed the depth lead, the surgeon went to remove the stylet from the depth lead and it sheared off right at the plastic clasp.The remainder of the stylet could not be removed from the depth lead and the surgeon had to explant the implanted depth lead.A replacement lead was implanted and the patient is currently implanted with the rns system programmed for detection.Additional follow up: the surgeon was set up for depth lead insertion when the stylet broke so there was no significant delay to reinsert the new lead (the fce estimated 10 minutes).The fce reported the patient was not affected.The lead was removed and a new depth lead was inserted during surgery.
|
| |
|