Model Number PM3242
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/29/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Analysis description: final analysis found excessive medical adhesive on the setscrews.The torque wrench could not be inserted and tighten the setscrew.The damage found likely resulted in the reported inability to tighten the setscrew.
 
Event Description
It was reported that during a device upgrade procedure, the left ventricular setscrew of the pulse generator was unable to be tightened.The device was not used and a new device was implanted successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE QUADRA RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
15900 valley view court
sylmar, CA 91342
8184934022
MDR Report Key5447947
MDR Text Key38587649
Report Number2017865-2016-01022
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2017
Device Model NumberPM3242
Device Lot NumberA000014413
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/29/2016
Initial Date FDA Received02/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial