Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a power (intermittent power) issue.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
The battery compartment was not damaged.The battery cap was not returned and a test battery cap was used and able to fit securely to the pump.The review of the black box data showed unexpected power interruptions on the date of the alleged issue occurrence.A call service alarm 69 was reproduced during the pump power up, indicating a language file corruption and a u39 component failure.The pump was unable to recover from the call service alarm 69 after a reboot attempt; therefore, the investigators were unable to adequately investigate the reported intermittent power complaint.Upon removal of the pump cover, no intermittent condition was found to the cgm (continuous glucose monitoring) module or to the printed circuit board.(b)(4).
|
|
Manufacturer Narrative
|
(b)(4).
|
|