Model Number 16-02-85
Device Problem Failure to Power Up (1476)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/21/2016
Event Type  No Answer Provided  
Manufacturer Narrative
Patient information was not provided.Sorin group (b)(4) manufactures the sorin heater-cooler system 3t.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group received a report that the sorin heater-cooler system 3t unit shut off by itself prior to a procedure, immediately after the unit was turned on.After multiple attempts to start the unit, the perfusionist was able to start the unit and it remained on for the length of the case.There was no report of patient injury.A sorin group field service representative was dispatched the facility and was unable to duplicate the issue.A replacement mains power switch was provided to the customer and the replaced device has been requested further investigation.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.User will return when replacement recv'd.
 
Event Description
Sorin group (b)(4) received a report that the sorin heater-cooler system 3t unit shut off by itself prior to a procedure, immediately after the unit was turned on.After multiple attempts to start the unit, the perfusionist was able to start the unit and it remained on for the length of the case.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key5452667
MDR Text Key38748568
Report Number9611109-2016-00113
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 01/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number16-02-85
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-