Model Number N/A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/02/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6).(b)(4).Investigation- the product was not returned to assist with this investigation; however, a review of the device history, quality control (qc), and instructions for use (ifu) was conducted for the purpose of this investigation.Per quality control specification, the tethering is inspected, verifying proper placement of suture within the stent; proper type of suture; proper handle, location and length; a secure knot and ensuring the tether is not tangled around the stent.A review of the device lot records and non-conformance history did not observe any quality related concerns associated with this case.The ifu details: "the stent may be removed easily by gentle withdrawal traction using endoscopic forceps." based on the information provided, the root cause is undetermined.Additional action is not required at this time, the residual risk remains at an acceptable level.We will continue to monitor for similar complaints.
 
Event or Problem Description
A ureteroscopy with double j stent placement was being performed on the male patient.The stent was placed on (b)(6) 2016.The tether was left on the stent and the physician instructed the patient to remove stent using the tether after 48 hours.Sometime within the 48 hour period, the tether broke and was not found in the patient (patient voided tether).Stent was removed via cystoscopy by the implanting physician.A section of the device did not remain inside the patient's body.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA FIRM URETERAL STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key5454176
Report Number1820334-2016-00098
Device Sequence Number19019706
Product Code FAD
UDI-Device Identifier10827002234052
UDI-Public(01)10827002234052(17)180803(10)6090238
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K961446
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberUFH-624-RT1
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/02/2016
Device Age6 MO
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2016
Initial Report FDA Received Date02/23/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-