Manufacturing review: the lot number has been provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported for this lot number and issue to date.Visual inspection: the sample was returned.The safety clip was missing upon receipt.The tuohy borst valve was tight, and the two-way stop cock was opened.The outer catheter was broken approximately 56mm from the strain relief and stretching was noted around the break.The stent graft was partially deployed and protruded approximately 1.5mm from the tip of the outer catheter.Bent struts were noted in the undeployed portion of the stent graft.No other anomalies were noted to the stent graft.Functional/performance evaluation: functional testing could not be performed, as the delivery system was returned with the outer catheter broken.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the investigation is confirmed for partial deployment, as the stent graft was returned partially deployed.It should be noted that an outer catheter break is also confirmed.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) state: indication for use: flair endovascular stent graft is indicated for use in the treatment at the venous anastomosis of eptfe or other synthetic arteriovenous (av) access grafts.Additionally, the current flair endovascular stent graft instructions for use (ifu) provides general instructions for use of the device, as well as warnings, precautions, and potential complications associated with the device.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|