Catalog Number 110940 |
Device Problems
Smoking (1585); Device Inoperable (1663); Temperature Problem (3022)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/08/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
|
Event Description
|
The surgeon was performing a partial knee arthroplasty using the robotic arm interactive orthopedic system (rio).During the case the anspach burr motor malfunctioned.The anspach motor was exchanged with a replacement anspach motor, which then also malfunctioned.The case was then completed successfully with a 3 minute delay.After the case when testing, one motor began smoking.
|
|
Manufacturer Narrative
|
Reported event: anspach emax 2 plus burr motor smoking during testing.Method & results: device evaluation and results: no device inspection could be completed as the device was not returned for evaluation.Device history review: not performed as the anspach emax 2 plus burr motor is an oem product.Complaint history review: based on the device identification, the catsweb and trackwise complaint databases were reviewed from 2011 to present for similar reported events regarding anspach emax 2 plus burr motor smoking during testing failure of p/n: 110940, sn: (b)(4).There have been no other similar events for the referenced serial number.Trend request for this part number has been submitted (trend request #737).Conclusions: the anspach emax 2 plus burr motor is an oem device.The failure mode could not be confirmed because the device was not returned for evaluation.Corrective action/preventive action: as the event did not involve a manufacturing related product problem indicating a non-conformity, adverse trend, or unanticipated hazard, no corrective action is required at this time.Trend request #737 has been initiated to continue to monitor for events related to this device.
|
|
Event Description
|
The surgeon was performing a partial knee arthroplasty using the robotic arm interactive orthopedic system (rio).During the case the anspach burr motor malfunctioned.The anspach motor was exchanged with a replacement anspach motor, which then also malfunctioned.The case was then completed successfully with a 3 minute delay.After the case when testing, one motor began smoking.
|
|