| Catalog Number |
482317550 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/27/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Per reported event; the polyaxial tulip head broke off of the screw while inserting the blocker.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: risk assessment: results: review of the correspondence confirms that the tulip disengagement occurred on the screw during final tightening.The sales representative confirmed that the torque wrench and anti-torque key were used as directed, however since the product is not available for evaluation.Conclusion: the cause of this issue is likely to be multi-factorial in nature and cannot be conclusively determined.
|
| |
|
Event or Problem Description
|
|
Per reported event; the polyaxial tulip head broke off of the screw while inserting the blocker.
|
| |
|
Search Alerts/Recalls
|