SYNTHES OBERDORF COMPACT AIR DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 511.701 |
Device Problem
Failure to Shut Off (2939)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/27/2013 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during service and repair, it was observed that the compact air drive device kept running during sawing.This event was not related to surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|