• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100H
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Electric Shock (2554)
Event Date 02/14/2016
Event Type  Injury  
Event Description
The recipient reportedly experienced shocking sensations with device use.External equipment was exchanged, however, the issue was not resolved.The recipient's device was explanted.
 
Manufacturer Narrative
The external visual inspection revealed the electrode was cut prior to receipt.This is believed to have occurred during revision surgery.In addition, the epoxy header was observed damaged.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed all of the electrical tests performed.The device failed the residual gas analysis test.The internal visual inspection revealed cracked conductive epoxy joints across multiple channels.The impedance measurements across these connections did not reveal any increased impedance.The failure of this device is attributed to excessive moisture through a gross leak at the case-band braze.This ultimately caused this device to cease functioning.This older device configuration is no longer manufactured.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
nicole birch
28515 westinghouse place
valencia, CA 91355
6613621528
MDR Report Key5470624
MDR Text Key39366149
Report Number3006556115-2016-00079
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberAB-5100H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2016
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age17 YR
-
-