PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V; JCX Back to Search Results
Catalog Number PMX110
Device Problems Device Emits Odor (1425); Smoking (1585); Device Operational Issue (2914); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/30/2016
Event Type  malfunction  
Manufacturer Narrative
Results: there was no visible damage to the exterior of the penumbra system aspiration pump max 110v (pump max).Conclusions: evaluation of the retuned device revealed that the pump was functional.The pump was powered on multiple times and the pump generated vacuum each time it was powered on.Furthermore, no unusual scents or sounds were detected coming from the pump.Therefore, the root cause of the complaint cannot be determined.The pumps are 100% functional tested during incoming inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110v (pump max).During the procedure, the pump max started smoking, making noises and smelling.The procedure was completed using the same pump max; however, the physician had to turn the pump max off and on multiple times throughout the procedure.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer (Section G)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
kathleen kidd
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key5471004
MDR Text Key39422907
Report Number3005168196-2016-00259
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue NumberPMX110
Device Lot NumberF17857-15
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-