The reason for this revision surgery was is the result of the patient shoulder fracture after 2.3 years of patient use.An infection was suspected but was not confirmed.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.Initial or prolonged hospitalization was required.The device was disposed of at the hospital and not made available to (b)(4) for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing history of this part.All parts were found to meet design and manufacturing specifications.No non-conforming material reports (ncmr's) were associated with this part.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.The sterility lot was reviewed and this lot was found to be incompliance with the specified sterility standards.This event is deemed as non-product related.The surgeon reported no issues associated with the explanted product and provided no information that definitively described the root cause or reason of the fracture.The scope of this investigation is limited without having the parts available for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
|