• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: S.U.A. MARTIN GMBH &CO.KG ULTRA-COAT KERRISON MICRO 40D UP 8 3MM; RONGEURS/KERRISONS/IVDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

S.U.A. MARTIN GMBH &CO.KG ULTRA-COAT KERRISON MICRO 40D UP 8 3MM; RONGEURS/KERRISONS/IVDS Back to Search Results
Catalog Number RB5863 / 130814BZ
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2013
Event Type  malfunction  
Manufacturer Narrative
To date the device involved in the reported incident has not been received for evaluation.It was not possible to review manufacturing records, since the lot # was unavailable.Since the device was not available for testing and evaluation, the cause of the problem cannot be established.(b)(4).
 
Event Description
Surgeon was performing a right l5-s1 microdiscectomy.Surgeon took a regular bite of bone from subarticular recess and device broke.Patient was not harmed.X-ray was not taken because broken piece was removed via microscope and matched with other piece to make sure.Customer reports device is "very old", at least 8 years old.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA-COAT KERRISON MICRO 40D UP 8 3MM
Type of Device
RONGEURS/KERRISONS/IVDS
Manufacturer (Section D)
S.U.A. MARTIN GMBH &CO.KG
uhlandstrasse 17
rietheim-weilheim, 78604
GM  78604
Manufacturer (Section G)
S.U.A. MARTIN GMBH &CO.KG
uhlandstrasse 17
rietheim-weilheim, 78604
GM   78604
Manufacturer Contact
markus pauli
uhlandstrasse 17
rietheim-weilheim, 78604
GM   78604
4619668540
MDR Report Key5472566
MDR Text Key39423694
Report Number8010386-2016-00004
Device Sequence Number1
Product Code HAE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberRB5863 / 130814BZ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age48 YR
-
-