| Catalog Number |
48551000 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Injury (2348); No Code Available (3191)
|
| Date of Event |
02/08/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the surgeon found the loosening of the blocker upon x-ray images.Therefore, the surgeon is monitoring the patient, and is planning revision surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Catalog# 48551000, lot# ln9.Method: visual inspection; functional inspection; device history review; complaint history review; risk assessment; results: manufacturing files were reviewed and no incidents were found.The returned blockers showed evidence of contact with a rod.Conclusion: the likely root cause is not determined due to lack of details available.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon found the loosening of the blocker upon x-ray images.Therefore, the surgeon is monitoring the patient, and is planning revision surgery.
|
| |
|
Search Alerts/Recalls
|