| Catalog Number |
2101-0005 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260); Material Integrity Problem (2978)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
| Date of Event |
02/11/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the implant cartridge was inserted into gun and the cartridge split and cracked when the tech was attempting to fill the cement cannula.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection; device history review; complaint history review; risk assessment; results: visual inspection indicated cortoss cartridge fractured at the proximal end this likely indicates an excessive force was applied through the delivery gun during application.Storage and room temperature, time allow for cortoss to equilibrate to room temperature, time and technique used to dispense the cortoss and/or the force used during application could be contributing factors for the reported event.Conclusion: the likely root cause is multifactorial.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant cartridge was inserted into gun and the cartridge split and cracked when the tech was attempting to fill the cement cannula.
|
| |
|