Results: the handle on top of the penumbra system aspiration pump max 110v (pump max) was loose on the left side when viewed from the front.Conclusions: evaluation of the returned device revealed that upon powering on, the pump could reach full vacuum pressure within specification, and was able to run without issue for 30 minutes.If the pump is turned off, the pressure inside the pump should be allowed to equalize before turning it back on.The root cause of the pump shutting off could not be determined.The pump functioned as intended upon evaluation.Further evaluation revealed that the handle on top of the pump was loose on one side.This failure was unassociated with the reported complaint.These devices are 100% functionally tested during inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110v (pump max).During the procedure, while aspirating clot, the pump max suddenly stopped on its own.After repeated attempts to turn on the pump max, it finally powered on within 20 seconds.The procedure was completed using this same pump max.There was no report of an adverse effect to the patient.
|