It was reported that high impedance (dcdc 7) was observed on vns patient's system.Normal mode and system mode diagnostics were run and high impedance were found (dcdc 7).It was reported that the device was then disabled.X-rays were taken and sent to the manufacturer for review.The generator appears in the upper left chest in a normal placement.The filter feed-through wires appeared to be intact.The lead connector pin is fully inserted into the generator connector block.The electrode appears to be placed in abnormal arrangement (the negative electrode appears to be placed distal to head) but this could not be confirmed due to poor quality of the image.It's unable to assess if the strain-relief bend and loop are presents.No tie-down was found holding the lead.A portion of lead behind the generator could not be assessed.No suspected lead break was found on the visible part of the lead.Further information was received from the physician, indicating that the patient's generator was implanted in 2014.It's unknown if any patient trauma or manipulation occurred that could be the cause of the suspected lead fracture.As reported by the parents to the physician, the patient's seizures were stable until (b)(6) 2016, without explanation.The last device checking in (b)(6) 2015 found that the parameters / settings were normal.During a clinic visit on (b)(6) 2016, high impedance was found.The device was then switched off that day.It was reported that lead revision is planned.No known surgical interventions have been performed to date.
|