Model Number N/A
Medical Device Problem Codes Fluid/Blood Leak (1250); Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Strain relief (rapidport ez).Medwatch sent to fda on 03/10/2016.The device was returned to apollo for analysis.A visual analysis was performed on the returned lap-band and noted the device was discolored, and it appeared blue in color.Particles were noted on the inner surface of the band.An air leak test was performed and noted leakage at the junction of the buckle and shell.A microscopic analysis was then performed on the device, and noted an unidentified opening between the shell and the buckle.Device labeling addresses possible outcomes of band restriction issues and leak(s) as follows: adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Unplanned deflation of the band may occur due to leakage from the band, the port or the connecting tubing.
 
Event or Problem Description
Reported as: a patient with the lap-band system was reported to have come "in due to no restriction.The reporter was unsure if any diagnostics tests were done prior to surgery.Surgery was done to evaluate if patient needed a revision.It was discovered that the lap-band was leaking (a hole) either between the buckle and the ring or the belt and the inner shell.The lap-band system was replaced.
 
Additional Manufacturer Narrative
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND AP STANDARD W/RAPIDPORT EZ
Common Device Name
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. captial of texas hwy
bldg 1, ste. 300
austin TX 78746
MDR Report Key5493698
Report Number3006722112-2016-00061
Device Sequence Number13521243
Product Code LTI
UDI-Device Identifier10811955020268
UDI-Public10811955020268
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P000008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2016
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/31/2016
Device Model NumberN/A
Device Catalogue NumberB-2360
Device Lot Number2688174
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/11/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/10/2016
Supplement Report FDA Received Date04/29/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
-
-