It was reported that during use of the device for a cardiopulmonary bypass procedure, the roller pump stopped.The device was not changed out, as the roller pump was able to be re-started and the case was completed successfully.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per the clinical review on 02-mar-2016: according to the supply chain (sc) associate, the case occurred on (b)(6) 2016.According to the sc associate, the roller pump stop was not related to a pump malfunction, but was caused by a faulty power source in the operating room (o/r).The pump was able to be re-started and the case was completed successfully.The pump was tested by the hospital biomed team after the procedure with a new power source and the pump worked without issue.There was no delay in the procedure and no associated blood loss.There was no harm observed.
|