• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751765
Device Problem Air Leak (1008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/25/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A nurse reported that an air leakage occurred with all three trocars during a vitreoretinal procedure.The procedure was completed after replacing the trocars.There was no patient harm.The samples are not available for evaluation.Additional information has been requested.
 
Manufacturer Narrative
No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.Due to an observed increased trend for this event, an internal investigation was initiated to determine if corrective and preventative actions were necessary.A root cause for the increased complaint rate was found to be related to a manufacturing post assembly inspection practice.This complaint does not identify a reported lot and no sample was returned for evaluation so it cannot be determined if the complaint can be attributed to the post assembly inspection.The post assembly inspection has been discontinued and an effectiveness check has been established and will be monitored periodically.A non-safety medical device correction was completed on august 12, 2015 and a manufacturing change implemented to address the root cause.All potentially impacted customers have been contacted, trocar plugs made available, and other risk mitigations discussed.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
eddie darton, md, jd
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686660
MDR Report Key5501920
MDR Text Key40634075
Report Number1644019-2016-00535
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeUP
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Remedial Action Notification
Type of Report Initial,Followup
Report Date 06/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Catalogue Number8065751765
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/24/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-