During removal, the angiosculpt separated in three pieces.Surgical intervention was required to remove the separated balloon from the patient resulting in prolongation of the procedure.Patient information regarding relevant tests/laboratory data or medical history is unknown.This information was not available from the facility.(b)(4).The angiosculpt was returned in three pieces (balloon, transition tubing, and shaft).Visual examination found the scoring elements struts slightly lifted from the distal bond.The intermediate bond was damaged and the scoring element rings were exposed.The transition tubing was necked, stretched, and separated from the intermediate bond and the proximal bond.In addition, the distal shaft was necked, stretched, and separated from the catheter shaft.The proximal bond was damaged and the inner member broke proximal to the proximal marker band which was located at the shaft.Based on the complaint information and the lab analysis, it is probable that the excessive force applied by the user contributed to the device separation.Per the ifu, retained device component is listed as a possible adverse effect of the procedure.
|
Prior to the use of the angiosculpt, ivus performed on the subclavian vein to treat patient with post thoracic outlet syndrome.The angiosculpt was loaded over a 0.014" ironman guide wire and introduced through 9 french, 10 cm terumo introducer sheath through the brachial vein.The angiosculpt was inflated three times at 12 atm for an average of 30 seconds each.The angiosculpt was removed and a venogram showed thrombus retained in the axillary vein.The angiosculpt was then reinserted and inflated four additional times at 30 seconds each.During removal, the balloon was unable to deflate and could not pass through the introducer sheath.Excessive force was applied to remove through the introducer sheath but the balloon detached from the shaft and was left in the brachial vein where it was surgically removed thru brachial cut down.
|