BAXTER HEALTHCARE - CALI MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number ARM4466E
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received for evaluation.A visual inspection was performed and identified the iodine sponge to be outside of the cap. a batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.The reported issue was verified; however, the cause could not be determined.There is a capa open to investigate this further.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the iodine sponge of a minicap was found outside of the minicap when the packaging was opened.This was noted before use.There was no patient injury or medical intervention reported.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE - CALI
apartado aereo 2446 calle #36,
cali
CO 
Manufacturer (Section G)
BAXTER HEALTHCARE - CALI
apartado aereo 2446 calle #36,
cali
CO  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5508415
MDR Text Key40634155
Report Number1416980-2016-05446
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeCO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 03/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2017
Device Catalogue NumberARM4466E
Device Lot NumberSE15JI1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-