DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.010
Device Problems Sticking (1597); Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/22/2016
Event Type  malfunction  
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was observed that the device did not run, the triggers were sticking and the coupling head was worn.It was further determined that the electric motor was damaged/corrosion, the bearings (rough rotation), the seals were damaged and there was corrosion on the internal gear components.It was further determined that the trigger components were damaged.The assignable root cause was determined to be due to component wear.If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that there was a possible short in the small battery drive device and the device ran intermittently.The reporter indicated that the event was discovered by a central sterile personnel.During in-house engineering evaluation, it was observed that the triggers were sticking.The event was not reported to have occurred during surgery.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5510880
MDR Text Key40863146
Report Number8030965-2016-11600
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 01/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.010
Device Lot Number003056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2016
Initial Date Manufacturer Received 03/08/2016
Initial Date FDA Received03/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-