• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. 3DKNEE; INSERT, 3D SIZE10 X 9MM L DCM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. 3DKNEE; INSERT, 3D SIZE10 X 9MM L DCM Back to Search Results
Catalog Number 391-09-610
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/23/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to poly wear and delamination of the medial compartment.
 
Additional Manufacturer Narrative
The reason for this revision surgery was wear and the medial compartment exhibited delamination after 6.3 years of patient.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to (b)(4) for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing history of this part.All parts were found to meet design and manufacturing specifications.No non-conforming material reports (ncmr's) were associated with this part.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.No information was provided that was descriptive or offered an explanation or extent of the delamination of the medial compartment.The surgeon reported no other issues associated with the explanted product.The scope of this investigation is limited without having the parts available for evaluation.Other conditions, root cause, relating to this event could not be determined with confidence.The wear and delamination is most likely the result extended use in the patient.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DKNEE
Common Device Name
INSERT, 3D SIZE10 X 9MM L DCM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key5515938
Report Number1644408-2016-00131
Device Sequence Number1096170
Product Code JWH
UDI-Device Identifier00888912118903
UDI-Public(01)00888912118903
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K020114
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2014
Device Catalogue Number391-09-610
Device Lot Number54000792
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/24/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/22/2016
Supplement Report FDA Received Date09/13/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-