| Catalog Number |
800804 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/01/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tubing on the inserter got cut by the inserter while inserting the implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Catalog# 800804.Lot# 06291501.Method: visual inspection; functional inspection; device history review; complaint history review; risk assessment; results: the inserter was returned for analysis and no issues were found therefore the reported event cannot be confirmed.The device was checked visually and functionally and it is fully functional.Conclusion: a plausible root cause could not be identified because the returned tl inserter was found to be fully functional.
|
| |
|
Event or Problem Description
|
|
It was reported that the tubing on the inserter got cut by the inserter while inserting the implant.
|
| |
|