| Catalog Number |
4430146 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/31/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch history review: the manufacturing file was reviewed.It is compliant with our specifications and no abnormality was detected.No other complaint was reported on this batch of access ports.Investigation: the involved device was discarded and not available for evaluation.X-ray examination: the review of x-rays picture taken just after implantation ((b)(6) 2015) shows that the catheter does not appear to have been correctly positioned on the port exit cannula during the implantation procedure.Conclusion: the disconnection of the catheter, observed only 1 month after the device implantation, is probably due to incorrect connection of the catheter to the port by user.No corrective action is envisaged.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2015: implantation of the access port system without difficulties.On (b)(6) 2015: catheter disconnected from the port.
|
| |
|
Additional Manufacturer Narrative
|
|
The additional information concerning the patient, received after the first medwatch report, does not changed the conclusion of the investigation.
|
| |
|
Search Alerts/Recalls
|