(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The steerable guiding catheter was received.The investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
This is filed based on returned device analysis that found the hemostasis valve bond was broken, which has the potential to cause a leak resulting in air being introduced into the patient.It was reported that this was a mitraclip procedure to treat mitral regurgitation (mr) with a grade of 3.The clip delivery system (cds) was attempted to be inserted into the steerable guiding catheter (sgc); however, the cds could not be inserted past the sgc keyway.The blue alignment lines were confirmed to be aligned.Three attempts were made and then the sgc was removed and replaced.The same cds was used in the new sgc without issue.It was confirmed that all steps were performed per the instructions for use during preparation of the devices and during the procedure.One clip was implanted, reducing the mr to 2.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.Subsequent, preliminary analysis of the returned device found that the sgc hemostasis valve bond was broken.
|