The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110v (pump max).During the procedure, the physician advanced a penumbra system 5max ace reperfusion catheter (5max ace) into the patient and switched the pump max on in order to start aspirating thrombus.However, the physician could not confirm that any vacuum was being produced and turned off the pump max.After verifying that the tubing was properly connected, the physician restored the regulator knob and turned the pump max back on but no vacuum was produced.Therefore, the procedure was completed using a stent.There was no report of an adverse effect to the patient.
|
Results: there was no visible damage to the pump.The pump was plugged in and powered on.The pump only made a humming sound like a fan was running, and no vacuum was produced.The pump was opened by penumbra engineers and corrosion was observed on the piston crown in the outlet cylinder.Conclusions: evaluation of the returned device revealed the pump was nonfunctional.The pump was opened by penumbra engineers and corrosion was observed on the piston crown in the outlet cylinder.The observed corrosion likely caused the piston to seize inside the cylinder.The root cause of this corrosion could not be determined.Penumbra pumps are 100% functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|