|
|
| Catalog Number |
520-46-024 |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
02/24/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision surgery - due to the subscapularis repair having failed; and causing rotator cuff deterioration.The surgeon wanted to convert to a reverse shoulder prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for this revision surgery was subscapularis repair failure causing rotator cuff deterioration.The length of the in-vivo service was 1.7 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to (b)(4) for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing history of this part.All parts were found to meet design and manufacturing specifications.No non-conforming material reports (ncmr's) were associated with this part.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed to be non-product related.The root cause for subscapularis repair failure was not reported; the surgeon reported no issues associated with the explanted product.The scope of this investigation is limited without having the parts available to (b)(4) for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
|
| |
|
Search Alerts/Recalls
|
|
|