| Catalog Number |
0066451 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Immucor technical support use a remote electronic connection method to access and assess the test wells on (b)(6) 2016.The test well in question visually appeared negative, which was consistent with the instrument output.The immucor laboratory tested retention product on (b)(6) 2016, which performed as expected.An immucor field service engineer (fse) visited the customer site on (b)(6) 2016 to assess the testing instrument and found that the instrument was operating as expected.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016, a customer reported an unexpected negative result when using gamma-clone anti-k (monoclonal) with galileo neo testing methodology, when tested on (b)(6) 2016.
|
| |
|
Search Alerts/Recalls
|