• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problems Break (1069); Crack (1135); Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a power (damage) issue.It was reported that the pump was experiencing an intermittent power issue.Reportedly, the battery compartment was cracked and the battery cap was unable to tighten on to the pump.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
 
Manufacturer Narrative
Follow-up #1 date of submission 04/26/2016 device evaluation: the pump has been returned and evaluated by product analysis on (b)(4) 2016 with the following findings: a review of the black box data showed unexplained reboots on (b)(4) 2016.A visual inspection of the pump showed that the battery compartment was cracked with stripped threads.Evidence of moisture was observed in the battery compartment had stripped threads.No battery cap was returned with the pump.A test cap was used and was unable to fully tighten on to the pump.The pump was able to power on with the test cap; the pump was exercised for 24 hours with no issues.The pump failed a leak test at the battery compartment.The pump cover was opened and no internal defect or moisture was found.The complaint was not duplicated during testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5547143
MDR Text Key42427517
Report Number2531779-2016-07177
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 03/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Other Device ID Number011084040610005121536739816
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age12 MO
Date Manufacturer Received03/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age45 YR
Patient Weight125
-
-