• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM; MDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM; MDS Back to Search Results
Model Number MT22495-BLU
Device Problem No Audible Alarm (1019)
Patient Problem Pneumonia (2011)
Event Date 02/03/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
Patient contacted dexcom on 03/08/2016 to report an adverse event that occurred on (b)(6) 2016.The patient was admitted to the hospital for pneumonia, where she stayed for 5 days and was given antibiotics.The doctor who treated her said that she had contracted pneumonia because her blood glucose levels were not under control at night and her receiver was not beeping when she went high/low.Additionally, the patient tested the audio function of the receiver and it did not beep.No further event or patient information was provided.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint device was returned for evaluation.The device was externally visually inspected and no defect was found.A review of the downloaded receiver log did not find any issues related to the customer complaint.Functional testing was performed and resulted in audio failures.The reported event of no audio output was confirmed.The root cause was determined to be a defective speaker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Type of Device
MDS
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key5549242
MDR Text Key41936835
Report Number3004753838-2016-02140
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000170
UDI-Public(01)00386270000170(241)MT22495-BLU(10)5200910(17)NA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Remedial Action Notification
Type of Report Initial,Followup
Report Date 03/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495-BLU
Device Catalogue NumberSTR-DR-BLU
Device Lot Number5200910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/08/2016
Initial Date FDA Received04/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age37 YR
-
-