| Catalog Number |
48230141 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Injury (2348); No Code Available (3191)
|
| Date of Event |
02/11/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the surgeon at the clinic, reported that he had to proceed to a revision surgery one month after the initial implantation because the external rod slipped from the junction connector and parted from it.Both sides of the assembly were concerned.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: risk assesment.Result: the reported event was confirmed via the correspondence with the international representative.However, the device was discarded and is unavailable for evaluation.No lot # was provided, so a manufacturing record review could not be performed.No device was received back, so testing and inspection could not be performed to aid in determination of the cause of the event.Conclusion: over or under tightening of the setscrews for this device can lead to loosening of the construct and possible disengagement of a rod attached to this device.However, because the device is unavailable for evaluation, this cannot be determined conclusively.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon at the clinic, reported that he had to proceed to a revision surgery one month after the initial implantation because the external rod slipped from the junction connector and parted from it.Both sides of the assembly were concerned.
|
| |
|