Model Number 2090
Device Problem Bent (1059)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/15/2016
Event Type  malfunction  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: manufacturer's analysis confirmed the customer comment that the radio frequency (rf) programmer head connector port on the programmer was loose.It was also indicated that programmer's lower display hinges were worn out, the sliding keyboard cover was missing, the right keyboard hinge was broken, and the hard drive was worn out.The programmer was to be scrapped.(b)(4).
 
Event Description
It was reported that the radio frequency (rf) programmer head connector port on the programmer was seemingly bent.Attempts with other rf heads showed no improvement.The programmer has been returned for repair.The programmer tested out of specification during manufacturer's analysis.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5552961
MDR Text Key42437502
Report Number2182208-2016-00790
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/15/2016
Initial Date FDA Received04/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-