• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA YORK, PA INC. MLX 300W XENON LIGHTSOURCE; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA YORK, PA INC. MLX 300W XENON LIGHTSOURCE; N/A Back to Search Results
Catalog Number 00MLX
Device Problems Thermal Decomposition of Device (1071); Device Sensing Problem (2917)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/16/2016
Event Type  malfunction  
Manufacturer Narrative
To date the device involved in the reported incident has not been received for evaluation.An investigation has been initiated based on the reported information.
 
Event Description
Customer reports that unit got hot enough to burn a hole through 2 headlight modules, two separate patient procedures.They were using the 001380lx9 cables.On (b)(6) 2016 customer has not provided additional promised information.Two of 2 related complaints (other report#2523190-2016-00059).
 
Manufacturer Narrative
The 6/21/16 integra investigation completed.Method: failure analysis, device history evaluation.Results: failure analysis - lamp is discolored.Control board failed error code sequence and memory.A visual inspection found that the turret is melted at the wolf port and the four bumper feet were missing.The turret, control board and 4 bumper feet were replaced during the repair.Device history evaluation - nonconforming product report / nonconforming material report history: none.Variance authorization / deviation history: none.Engineering change order / manufacturing change order history: none.Corrective action preventive action history: none.Health hazard evaluation history: none.Conclusion: the complaint report is confirmed.The unit was repaired and returned to customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MLX 300W XENON LIGHTSOURCE
Type of Device
N/A
Manufacturer (Section D)
INTEGRA YORK, PA INC.
589 davies drive
589 davies drive
york PA 17402
Manufacturer (Section G)
INTEGRA YORK, PA INC.
589 davies drive
york PA 17402
Manufacturer Contact
sandra lee
311 enterprise drive
plainsboro, NJ 08536
6099362393
MDR Report Key5555719
MDR Text Key42160639
Report Number2523190-2016-00060
Device Sequence Number1
Product Code FCW
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00MLX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/25/2016
Initial Date FDA Received04/07/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/09/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-