Catalog Number B12SRT
Device Problems Detachment Of Device Component (1104); Component Falling (1105)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/15/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The following information was requested, but unavailable: can you please clarify if the outer seal you have referred to that fell on the floor when the package was opened, is the reducer/attachment cap (universal seal cap)? no information.
 
Event Description
It was reported that before a laparoscopic-assisted distal gastrectomy (ladg), the outer seal came off and fell on the floor when the package was opened.The device was not used for the patient.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Batch # = m93k5n.The analysis results found that the b12srt device was returned in good condition and out of its sterile package.In addition the universal seal was returned not latched to the sleeve.In an attempt to replicate the reported incident, the universal seal latch onto the sleeve assembly.In addition the obturator latched onto the universal seal as well.No seal issues were noted during functional testing.No conclusion could be reached as to what may have caused the reported incident.A batch record review was performed and no anomalies were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key5555827
MDR Text Key42747659
Report Number3005075853-2016-01937
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/17/2020
Device Catalogue NumberB12SRT
Device Lot NumberM4JA7E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/16/2016
Initial Date FDA Received04/07/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-