IPG MFG SWITZERLAND REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problems Premature Discharge of Battery (1057); False Positive Result (1227)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the battery impedance trend was rising.Analyst commented, recommended replacement time (rrt) alert was triggered on (b)(6) 2016.Daily battery trend data shows gradual rise/varying battery impedance.Premature rrt alert, without corresponding battery depletion.(b)(4).
 
Event Description
It was reported that during use the implantable cardiac monitor (icm) encountered an incident where an algorithm falsely triggered recommended replacement time (rrt).The icm remains in use as the devices will continue to function in spite of the false rrt indication.Upon eos trigger the icm will automatically switch to non-wireless telemetry only.The patient will be able to transmit manually (or non-wirelessly with the patient wand).No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH  1131
Manufacturer (Section G)
IPG MFG SWITZERLAND
route du molliau 31
tolochenaz 1131
CH   1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5559917
MDR Text Key42765696
Report Number9614453-2016-02030
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/14/2015
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2016
Initial Date FDA Received04/07/2016
Supplement Dates Manufacturer Received02/15/2016
Supplement Dates FDA Received09/20/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/15/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-