• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. FMP HIP; SHELL, ACETABULAR, FLARED FMP, NO HOLES SIZE56

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. FMP HIP; SHELL, ACETABULAR, FLARED FMP, NO HOLES SIZE56 Back to Search Results
Catalog Number 430-04-056
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Date of Event 03/11/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient complaining of pain and discomfort.The cup and liner failed; the had cup become dislodged from original position.
 
Additional Manufacturer Narrative
The reason for this revision surgery was pain and discomfort in the hip.The length of the in-vivo service was 11 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing history of this part.All parts were found to meet design and manufacturing specifications.No non-conforming material reports (ncmr's) were associated with this part.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed to be non-product related.The cup had become dislodged from the original position; no definitive information was provided that definitively described the root cause for the cup movement and dislodging.The surgeon reported no issues associated with the explanted product.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP HIP
Common Device Name
SHELL, ACETABULAR, FLARED FMP, NO HOLES SIZE56
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key5568848
Report Number1644408-2016-00198
Device Sequence Number462665
Product Code LPH
UDI-Device Identifier00888912082419
UDI-Public(01)00888912082419
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K974095
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2008
Device Catalogue Number430-04-056
Device Lot Number813761
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 03/14/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/11/2016
Supplement Report FDA Received Date06/13/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient Weight79
-
-