It was reported that the disposable device was discarded by the user and is not available to be returned to the manufacturer for evaluation.An investigation into the root cause of this incident is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.(b)(4).Device is unavailable for return.
|
|
As reported (b)(6) 2016, a patient of unknown age and gender presented for an endovenous laser procedure.During preparation for the procedure, when the laser fiber was removed from the sterile packaging, it was noted the fiber was fractured.The device was set aside and a new of the same device was used to successfully complete the procedure.There was no harm or injury to the patient due to the event as the device did not come into contact with the patient.It was reported the disposable device is not available for return to the manufacturer for a device evaluation as it was disposed of by the user.
|
|
As the device was not returned, angiodynamics is unable to perform a device evaluation.However, the customers reported complaint of a broken (cracked) fiber is confirmed based on the photographic evidence provided.Although the complaint description is confirmed, a definitive root cause for the event cannot be determined.Angiodynamics' supplier of this device was informed of the event.The supplier performed a review of the device history records for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".During the manufacturing process, the device goes through several aql inspections.Each unit is also repeatedly bent and coiled during the processing and there is a final visual inspection performed with a hene laser to check the entire fiber length for defects prior to packaging and shipment.During the hene laser test the fiber tip is examined closely for damage.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint #(b)(4).Device not returned, user photos provided.
|