Conmed corporation received one (1) "unopened" packages containing the 1/4" x 6' suction tubing for evaluation.Visual examination of the returned packages found the package with full seal disruption occurring in the final manufacturing seals.In the returned package, there is evidence that the device, suction tubing, was sealed within the sterile seal of the package.In that area adhesive transfer to the poly material showing that a full seal existed is missing due to the product obstructing the seal.The device was transferred to packaging engineering for evaluation and testing.This device was confirmed that the seal failed the dye leak test on 12-apr-2016.This lot was manufactured on 17-mar-2015.A review of the device history record for this lot found no ncrs and no note of discrepancies during the manufacturing process.In these instances, preliminary investigation revealed that the most probable cause of the seal disruptions appeared to be related to the device being within the sterile seal area during the final manufacturing sealing process.(b)(4).The reported packaging anomalies were obvious to the distributor and therefore prompted the return of the devices for evaluation and replacement.To date, there have been no serious injuries or death related to this reported problem.Nonetheless, to prevent future recurrences, an investigation has been opened to address this issue.These devices were manufactured prior to the opening of that investigation.As with all medical devices, examination of packaging occurs multiple times prior to use (shipping/receiving, distribution, storage and prior to use).In addition, good clinical practice would include examination and verification of the product and its original packaging to ensure both are intact.If the product and its packaging have been opened/damaged or altered, do not use the product and contact the manufacturer immediately.
|
|
The distributor in (b)(6) reported that during receiving and inspection of incoming products, one (1) package containing 1/4" x 6' suction tubing was discovered with "insufficient heat seal".The returned package exhibited full open seal disruption in the final manufacturing seals.In this instance, there was no patient involvement with this reported problem, as the packaging anomaly was discovered during inspection at the distributor facility prior to distribution to an end-user.
|