Boston scientific received information that this patient was presented to the hospital after receiving multiple appropriate shocks from their implantable cardioverter defibrillator (icd) due to ventricular tachycardia.Upon interrogation, fault code 01 was exhibited which is indicative of the shocking capacitors not being charged to the programmed voltage within 45 seconds after the start of charging.Due to the presence of the code, the icd also triggered end of life.No intervention was performed and the tachy therapy was programmed off and the icd will remain implanted and in service.No adverse patient effects were reported.
|