GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T167
Device Problems Defibrillation/Stimulation Problem (1573); Device Displays Incorrect Message (2591)
Patient Problem Ventricular Tachycardia (2132)
Event Date 12/18/2015
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this patient was presented to the hospital after receiving multiple appropriate shocks from their implantable cardioverter defibrillator (icd) due to ventricular tachycardia.Upon interrogation, fault code 01 was exhibited which is indicative of the shocking capacitors not being charged to the programmed voltage within 45 seconds after the start of charging.Due to the presence of the code, the icd also triggered end of life.No intervention was performed and the tachy therapy was programmed off and the icd will remain implanted and in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5593838
MDR Text Key43197759
Report Number2124215-2016-00629
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/05/2008
Device Model NumberT167
Other Device ID NumberVITALITY 2 DR EL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 01/01/2016
Initial Date FDA Received04/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4087
Patient Age96 YR
-
-