The case states that the rapicide pa part b high level disinfectant (hld) dosing chamber hoses were crossed on a medivators dsd edge automated endoscope reprocessor.Medivators field service engineers were on site the next day to repair the machine and it performs according to specification.The scopes reprocessed in this machine were appropriately disinfected as the part a component was dosed according to specification.Rapicide pa part a is the component in the hld that contains the active ingredients.There is potential chemical exposure to the rapicide pa part b component if scope was left sitting in the opposing basin.To date, there have been no reports of chemical exposure to the handler or patient.This complaint will continue to be monitored within medivators complaint system.
|