Model Number U128
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Patient Involvement (2645)
Event Date 12/23/2015
Event Type  malfunction  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.Analysis of device memory noted a leakage on lead fault message was recorded while the device was still in the box and in storage mode.The root cause of the fault was unable to be determined.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.
 
Event Description
(b)(4) received information that this product was returned from our final pack area for analysis as the product had not passed all required testing.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5596726
MDR Text Key43278161
Report Number2124215-2016-00393
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public(01)00802526559402(17)20170528
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/28/2017
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 01/06/2016
Initial Date FDA Received04/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-