| Catalog Number |
FEM14120 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/02/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide any patient details.
|
| |
|
Event or Problem Description
|
|
It was reported that during stent deployment, excessive deployment force was encountered resulting in the outer catheter of the delivery system to be separated.The device was removed and another stent graft was attempted to be deployed unsuccessfully.A third stent graft was used to complete the procedure successfully.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.The evaluation of the returned sample confirmed that the stent graft could not be deployed completely.The outer sheath was found to be torn off and elongated indicating that high force must have affected on the delivery system during the attempt to release the stent graft.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This type of event may be associated with a difficult vessel anatomy leading to increased friction and subsequent sheath fracture.In this case, no information regarding the vessel anatomy was provided.Insufficient flushing of the device may be another contributing factor to the reported event.In this case, no information regarding the preparation of the device was provided.On the basis of the information available, a definite root cause for the reported failure could not be determined.The ifu states that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure" and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
|
| |
|
Search Alerts/Recalls
|