Model Number R SERIES |
Device Problem
Poor Quality Image (1408)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a follow-up report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device's display showed vertical lines and was not legible.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing without duplicating the malfunction.The lcd display color cable was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
|