GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T175
Device Problem Communication or Transmission Problem (2896)
Patient Problem Syncope (1610)
Event Date 03/16/2016
Event Type  Injury  
Manufacturer Narrative
Additional information was received indicating the patient elected to not have the device replaced.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that the patient implanted with this implantable cardioverter defibrillator (icd) presented due to syncope.Attempts to interrogate this device were unsuccessful.A magnet was applied to the device and there was no evidence of tones.Boston scientific technical services (ts) discussed the device has been implanted for almost 10 years and the battery may be depleted.The patient reported they had their device checked two months ago with no anomalies noted.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5602954
MDR Text Key43465450
Report Number2124215-2016-05267
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/21/2007
Device Model NumberT175
Other Device ID NumberVITALITY 2 VR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 03/16/2016
Initial Date FDA Received04/25/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
T175
Patient Outcome(s) Life Threatening;
Patient Age61 YR
-
-