DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.110 |
Device Problems
Vibration (1674); Device Contamination with Chemical or Other Material (2944); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device was running rough with the electric motor vibrating and emitting an unusual noise.It was also reported that the components had an unknown substance/residue on them.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to failure to follow cleaning, sterilization and/or maintenance procedures and component wear.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the small battery drive device was making a grinding noise.During an in-house engineering evaluation, it was observed that the device was running rough with the electric motor vibrating and emitting an unusual noise.It was also reported that the components had an unknown substance/residue on them.It was not reported whether the device was used in a surgical procedure or if there was patient involvement.It was not reported if there were any delays in a scheduled surgical procedure or if a spare device was available for use.There were no reports of any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|